2.2 Area of Application
The product is recommended for mobility grade 2 (restricted outdoor walker) and
mobility grade 3 (unrestricted outdoor walker).
kg
Size [cm]
Max. body weight [kg]
2.3 Environmental Conditions
Allowable environmental conditions
Temperature range for use: -10 °C to +60°C
Allowable relative humidity 0 % to 90 %, non-condensing
Unallowable environmental conditions
Mechanical vibrations or impacts
Perspiration, urine, fresh water, salt water, acids
Dust, sand, highly hygroscopic particles (e.g. talcum)
2.4 Service Life
This product has been tested by the manufacturer with 2 million load cycles according to ISO
22675. Depending on the patient's activity level, this corresponds to a service life of 2 to 3 years.
3 Safety
3.1 Explanation of Warning Symbols
Warning regarding possible risks of accident or injury.
CAUTION
NOTICE
Warning regarding possible technical damage.
3.2 General Safety Instructions
CAUTION
Exceeding the service life and reuse on another patient
Fall due to loss of functionality as well as damage to the product
Ensure that the approved service life is not exceeded (see section "Service Life").
Only use the product for a single patient.
Inform the patient.
CAUTION
Excessive strain on the product
Fall due to breakage of load-bearing components
Use the prosthesis components according to the MOBIS classification (see section "Area of
Application").
Inform the patient.
10 | Ottobock
22 to 23
24 to 25
75
26 to 28
100
125