Simultaneous And Alternating Stimulation - Physiomed IONOSON-Basic Operating Instructions Manual

Mains- / battery-operated
Table of Contents

Advertisement

3 Overview of Current Modes
(FM) Frequency-Modulated Current
(STOCH) Stochastic Current
(T/R) Pulses with Adjustable Parameters
(IFM) Modulated Interference Current
(KOTS) Medium-Frequency Muscle Training

3.2.1 Simultaneous and Alternating Stimulation

In current mode FaS and KOTS you can increase and fade the currents in channel 1 and 2 either simultan-
eously or alternating. Three different modes of coordination are possible (simultaneous, directly altern-
ating and alternating with pauses):
How to Make a Treatment with a Preselected Pulse Duration and Pause
(1)
In the Display <1>, select a suitable current mode e.g. KOTS.
(2)
With the Data Selector <2>, click on the symbol
12
on page 27
on page 29
Automatic triggering: Channels 1 und 2 stimulate simultaneously with identical para-
meters.
Manual control (
tion in both channels until release.
Manual triggering (
stimulation in both channels with the length set in T.
Automatic triggering: Channels 1 and 2 provide alternating stimulation in immediate
succession. This is followed by a surge cycle consisting of:
Increase in channel 1 with the length selected for T
No pause
Increase in channel 2 with the length selected for T
Pause with the length set for R
Manual control (
channel 1 until release, second pressing causes stimulation in channel 2 until release.
Manual triggering (
in channel 1 with the length set in T. The corresponding pulse in channel 2 follows
automatically without a pause.
Automatic triggering: Channels 1 and 2 provide alternating stimulation, with pauses
defined between increases in channel 1 and 2.
Manual control (
channel 1 until release, second pressing causes stimulation in channel 2 until release.
Manual triggering (
in channel 1 with the length set in T, second pressing causes stimulation in channel 2
with the length set in T.
on page 32
on page 33
on page 36
): Pressing the manual release key causes simultaneous stimula-
): Pressing the manual release key causes simultaneous
): First pressing of manual push button causes stimulation in
): First pressing of manual push button causes stimulation
): First pressing of manual push button causes stimulation in
): First pressing of manual push button causes stimulation
in channel I. The following window opens: