5 Therapy with Stimulation Current
How to Perform a Treatment with Automatic Current Output
(1)
Switch on the instrument.
(2)
Attach the electrodes to the respective stimulation points. Position the patient to enable free move-
ment of the stimulation area.
(3)
Depending on the indication, use the Display <1> to select one of the current modes: FaS or T/R.
(4)
If necessary, select the biphase pulse emission.
(5)
Activate the acoustic signal if desired.
(6)
Set the pulse time T and the release time R. The release time R should be selected to give the nerve-
muscle-unit sufficient time for recreation.
(7)
Select the ramp of the current rise (for FaS) or the mode of the pulse (for T/R).
(8)
Set the therapy time.
(9)
If necessary, plug in the manual release key as therapy pause button in the Manual Release Key
Socket <11> on the front of the instrument and hand the manual release key to the patient.
(10) Slowly turn up the intensity of the respective channel (Intensity Control Channel I <3> or Intens-
ity Control Channel II <4>). The therapy time elapses.
CAUTION
Make sure you do not touch the electrodes when the current is turned up!
(11) Signs of fatigue should make you stop the treatment (they may already occur after a few muscle con-
tractions)!
(12) After expiry of the timer, an acoustic signal will be issued. The intensity is turned down to zero auto-
matically. Remove the electrodes.
How to Perform a FaS Treatment with Manual Release Key
Treatment with a manual release key is advisable for individual therapy methods in the field of sports, for
intentional exercises according to Förster as well as for practising both co-ordination and nerve tracts. This
is the great advantage of working with the manual release key.
(1)
Switch on the instrument.
(2)
Plug the manual release key into the Manual Release Key Socket <11> at the front of the instru-
ment.
(3)
Attach the electrodes to the respective stimulation points. Position the patient to enable free move-
ment of the stimulation area.
(4)
Depending on the indication, use the Display <1> to select the current mode FaS.
(5)
If necessary, select the biphase pulse emission.
(6)
Activate the acoustic signal if desired.
(7)
If necessary, set the stimulation intensity Ti, the stimulation frequency Ri or the pulse mode.
(8)
With the Data Selector <2>, click the symbol
Selection window for channel I
48
PHYSIOMED ELEKTROMEDIZIN AG
in channel I. The following window opens: