Physiomed IONOSON-Basic Operating Instructions Manual page 59

Mains- / battery-operated
Table of Contents

Advertisement

PHYSIOMED ELEKTROMEDIZIN AG
How to Make a Treatment with Medium Frequency Stimulation Current KOTS
(1)
Switch on the instrument.
(2)
Attach the electrodes to the respective stimulation points. Position the patient to enable free move-
ment of the stimulation area.
(3)
Select the current mode KOTS.
(4)
Set the pulse time T and the release time R. The release time R should be selected to give the nerve-
muscle-unit sufficient time for recreation.
(5)
Select the ramp of the current rise.
(6)
Select the modulation frequency.
(7)
If necessary, alter the basic frequency F
(8)
Activate the acoustic signal if desired.
(9)
Set the therapy time.
(10) If necessary, plug in the manual release key as therapy pause button in the Manual Release Key
Socket <11> on the front of the instrument and hand the manual release key to the patient.
(11) Slowly turn up the intensity of the respective channel (Intensity Control Channel I <3> or Intens-
ity Control Channel II <4>). The therapy time elapses.
CAUTION
Make sure you do not touch the electrodes when the current is turned up!
(12) Signs of fatigue should make you stop the treatment (they may already occur after a few muscle con-
tractions)!
(13) After expiry of the timer, an acoustic signal will be issued. The intensity is turned down to zero auto-
matically. Remove the electrodes.
How to Control a Treatment with KOTS with the Manual Release Key
(1)
Switch on the instrument.
(2)
Plug the manual release key into the Manual Release Key Socket <11> at the front of the instru-
ment.
(3)
Attach the electrodes to the respective stimulation points. Position the patient to enable free move-
ment of the stimulation area.
(4)
Select the current mode KOTS.
(5)
With the Data Selector <2> first click the R symbol, rotate to the left until one of the manual con-
trol symbols appears and press to confirm. The symbols have the following meaning::
releasing the manual release key.
(6)
With the Data Selector <2> click the symbol
lease key symbol is displayed next to the pulse symbol.
Important
If you use the manual release key, you can only turn the intensity up in both channels if
you press the manual release key (security blocking)!
(7)
Select the ramp of the current rise.
IONOSON-Basic
: The duration of increase is determined by pressing the manual release key, the pause by
: Pressing the key triggers a single pulse of the preset increase duration T.
.
bas
, to return to the KOTS menu. The manual re-
5 Therapy with Stimulation Current