Physiomed IONOSON-Basic Operating Instructions Manual page 34

Mains- / battery-operated
Table of Contents

Advertisement

3 Overview of Current Modes
Menu for Current Mode FM
Current Mode FM: monophase
Parameters for Current Type FM
Stimulation fre-
quency
Pulse time
Release time
How to Change the Pulse Shape
(1)
With the Data Selector <2>, click on the symbol
(2)
Turn the Data Selector <2> to the right by one position to make the pulse shape change and press
it for confirmation.
NOTICE
You can only change the pulse shape from triangle to square if the intensity in both channels
is zero. Square pulses make the current mode more intense.
28
Polarity switch to switch the polarity of the two electrodes ( the sign in the intensity dis-
play changes accordingly); can be adjusted separately for both channels.
Switching from monophase to biphase pulse emission; can be adjusted separately for
both channels.
Pulse shape (triangular pulse, square pulse); can be adjusted separately for both chan-
nels
7 - 14 Hz
1 ms
70 - 142 ms
Current Mode FM: biphase
. The symbol starts flashing.
PHYSIOMED ELEKTROMEDIZIN AG
IONOSON-Basic