Physiomed IONOSON-Basic Operating Instructions Manual page 41

Mains- / battery-operated
Table of Contents

Advertisement

PHYSIOMED ELEKTROMEDIZIN AG
Current Mode AMF: Sinusoidal Waves
Parameters for Current Mode AMF
Pulse shape
Relation between Frequency Cycle and Frequency Band
You can only select a frequency cycle if you have selected or programmed a frequency band
using the MODE function. Before this, you cannot click the symbol F
For the frequency cycle to become active, a frequency band has to be selected!
Example for a Frequency Cycle
Let's suppose you have selected the frequency cycle 1 s / 30 s / 1 s and the frequency band 0
- 250 Hz:
First, the frequency is on the maximum value (250 Hz) for one second. Afterwards, the fre-
quency is diminishing continuously to zero for 30 seconds. This frequency persists for one
second. At last the frequency is increasing continuously, until the upper limit frequency (250
Hz) is reached after 30 seconds, and the cycle restarts.
How to Program a Frequency Band
(1)
Click on the symbol
(2)
Turn the Data Selector <2> to the right until the symbol <<F>> is displayed and press to confirm.
(3)
Turn the Data Selector <2> to the right by one position until it reaches the lower barrier frequency
and press it.
(4)
Turn the Data Selector <2> to set the desired frequency and press it to confirm.
(5)
Turn the Data Selector <2> one position to the right to set the upper limit frequency and press to
confirm.
(6)
Turn the Data Selector <2> to set the desired frequency and press it to confirm.
IONOSON-Basic
Frequency cycle (1 s / 30 s / 1 s; 1 s / 1 s / 1 s; 3 s / 3 s / 3 s; 6 s / 6 s / 6 s; 12 s
/ 12 s / 12 s; Time upper frequency / Changing time / Time lower frequency)
Basic frequency (2.0 - 9.5 kHz). A lower frequency enhances the stimulation
intensity, a higher frequency lowers it.
Frequency bands (0 - 25 Hz; 0 - 50 Hz; 0 - 250 Hz; 100 - 250 Hz; free pro-
grammable 0 - 250 Hz; or fixed frequency SELECTIVE 0 - 250 Hz); the currently
active frequency is displayed to the right next to the symbol f.
Sinusoidal waves
by means of the Data Selector <2>. The symbol starts flashing.
3 Overview of Current Modes
!
cycl
35

Hide quick links: