3 Overview of Current Modes
Menu for Current Mode HV
Current Mode HV: monophase
Parameters for Current Mode HV
Pulse shape
How to Set the Pulse Time
(1)
Click on the Ti symbol using the Data Selector <2>. The symbol starts flashing.
(2)
Turn the Data Selector <2>, to adjust the desired pulse time and confirm by pressing it.
NOTICE
You can only extend the pulse time Ti if the intensity in both channels is zero. If you ex-
tend the pulse time, the current is experienced by the patient as more intense.
How to Adjust the Selective Frequency
(1)
Click on the symbol
(2)
Turn the Data Selector <2> to the right until the SELECTIVE symbol is displayed.
(3)
Turn the Data Selector <2> and turn it to the right by one position.
22
Polarity switch to switch the polarity of the two electrodes (the sign in the intensity dis-
play changes accordingly); can be adjusted separately for both channels.
Switching from monophase to biphase pulse emission.
Pulse time Ti (80 - 500 µs)
Burst frequency (1 - 10 Hz)
Frequency bands 1 - 150 Hz, 70 - 150 Hz or Selective frequency 1- 200 Hz (classical:
197 Hz)
Twin Pulses
by means of the Data Selector <2>. The symbol starts flashing.
Current Mode HV: biphase
PHYSIOMED ELEKTROMEDIZIN AG
IONOSON-Basic