Chapter 3 Overview of Current Modes

NOTICE
All current modes of the instrument are emitted in the constant current mode unless other-
wise specified (refer to
ment (refer to
a mixed mode consisting of constant current and constant voltage mode to avoid unpleasant
current sensations when the contact between skin and transducer is interrupted.

3.1 Settings for all Current Modes

The following settings are valid for all current modes:

3.2 Dual Channel Operation

Most current modes can be applied in dual channel operation. The following current modes support dual
channel operation:
IONOSON-Basic
Constant Current Operation
Simultaneous Therapy
Duration of therapy stated in minutes and seconds: The therapy duration of the re-
spective channel is preselected with the therapy timer.
Click on the corresponding function of the timer using the Data Selector <2> (the
symbol
flashes) and turn the Data Selector <2> to the right or left (possible val-
ues: 1.00 ... 60 min). The respective timer is started when increasing the intensity with
the Intensity Control Channel 1 <3> or the Intensity Control Channel 2 <4>.
If the intensity in a channel is turned down to zero, the timer assigned to the channel
stops. When the therapy time is over, you hear a beep. The intensity of the respective
channel is turned down slowly. During a treatment, the timer automatically sets the
therapy time used last.
Current intensity set: This field displays the intensity of the respective channel in mA
(milliampere) or µA (microampere; only for GMC and MENS current modes).
The bar graph below illustrates the maximum level of intensity you can set. The sign
next to the intensity displayed indicates the polarity of the electrode with the red con-
nector.
on page 25
on page 26
on page 40). Only in simultaneous treat-
on page 57) the instrument is automatically switched to
on page 23
3 Overview of Current Modes