Ig30) Pulse Galvanization 30; Ig30) Pulse Galvanization - Physiomed IONOSON-Basic Operating Instructions Manual

Mains- / battery-operated
Table of Contents

Advertisement

PHYSIOMED ELEKTROMEDIZIN AG
Current Mode MENS: monophase
Parameters for Current Mode MENS
Pulse time
Pulse shape
How to Adjust the Selective Frequency
(1)
Click on the symbol
(2)
Turn the Data Selector <2> to the right until the SELECTIVE symbol is displayed.
(3)
Turn the Data Selector <2> and turn it to the right by one position.
(4)
Press the Data Selector <2>. The arrow symbol starts flashing.
(5)
Press the Data Selector <2> and adjust the desired frequency (the desired frequency is displayed
next to the f symbol) and press it for confirmation.
NOTICE
If you select the selective frequency 10 Hz, you cannot use the burst function.
The maximum intensity of current mode GMC is so small that it is hardly perceptible. Its
energy is not sufficient to light Patient Current Indicator <5> and Output Indicator
<6>. The diodes therefore remain dark.

3.3.11 (IG30) Pulse Galvanization 30

Definition
Application
IONOSON-Basic
Polarity switch to switch the polarity of the two electrodes; the sign in the intensity dis-
play changes accordingly.
Switching from monophase to biphase pulse emission.
Burst frequency (0.1 - 15.0 Hz)
Frequency bands 50 - 200 Hz, 50 - 300 Hz, 200- 300 Hz or selective frequency 10 Hz or
50 - 300 Hz
1.6 ... 50 ms
Square pulse
by means of the Data Selector <2>. The symbol starts flashing.
Pulse Galvanization with strongly stimulating pulse frequency; acc. to Prof. Jantsch
Circulatory stimulation
Analgesia
Current Mode MENS: biphase
3 Overview of Current Modes
25