Physiomed IONOSON-Basic Operating Instructions Manual

Mains- / battery-operated
Table of Contents

Advertisement

O P E R A T I N G I N S T R U C T I O N S
IONOSON-Basic
M A I N S - O P E R A T E D M O D E L
B A T T E R Y - O P E R A T E D M O D E L
01750 GB

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the IONOSON-Basic and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Physiomed IONOSON-Basic

  • Page 1 O P E R A T I N G I N S T R U C T I O N S IONOSON-Basic M A I N S - O P E R A T E D M O D E L...
  • Page 2 Distribution, reproduction and translation of the software and its documentation (or excerpts thereof) are prohibited without the prior written consent of PHYSIOMED ELEKTROMEDIZIN AG. PHYSIOMED ELEKTROMEDIZIN AG reserves the right to change the software and associated data as well as the documentation without notice. All other rights reserved.
  • Page 3: Table Of Contents

    Table of Contents PHYSIOMED ELEKTROMEDIZIN AG Table of Contents Chapter 1 Introduction ........1 1.1 Conventions Used .
  • Page 4 Table of Contents PHYSIOMED ELEKTROMEDIZIN AG 3.3.6 (LP) Diadynamic Current Modulé en Longues Périodes ... . . 19 3.3.7 (UR) Ultra Stimulation Current......20 3.3.8 (HV) High-Voltage Current .
  • Page 5 Table of Contents PHYSIOMED ELEKTROMEDIZIN AG 5.5.4 Muscle Stimulation Using Low-Frequency Currents ....47 5.6 Therapy with Medium Frequency Current Modes....50 5.6.1 Treatment with IFM Current.
  • Page 6 Table of Contents PHYSIOMED ELEKTROMEDIZIN AG Appendix C Supplementary Documents ......74 C.1 Manufacturer’s Recommendations ..... . . 74 Index .
  • Page 7: Chapter 1 Introduction

    PHYSIOMED ELEKTROMEDIZIN AG Chapter 1 Introduction With IONOSON-Basic you have acquired a high-quality and extremely versatile combination therapy instru- ment. The instrument will only show its true potential, however, if you are well informed about its func- tions. For this reason, carefully read the Operating Instructions and familiarize yourself with the use of the instrument.
  • Page 8: Instrument Overview

    1 Introduction PHYSIOMED ELEKTROMEDIZIN AG 1.3 Instrument Overview Legend Indicator Mains Module Data Selector Power Switch Intensity Control Channel 1 Patient Lead Connector Intensity Control Channel 2 Ultrasound Lead Connector Patient Current Indicator Manual Release Key Socket Output Indicator IONOSON-Basic...
  • Page 9: General Notes

    PHYSIOMED ELEKTROMEDIZIN AG 1.4 General Notes IONOSON-Basic may only be operated in dry rooms designed for medical purposes, in accordance with DIN VDE 0100-710, User Class 1. These are rooms with patient beds, physical therapy rooms and practice rooms. It complies with the technical specifications of VDE 0750/IEC 60601 and is assigned to class IIa ac- cording to the Council Directive concerning Medical Devices.
  • Page 10: Application

    1 Introduction PHYSIOMED ELEKTROMEDIZIN AG The function of IONOSON-Basic are controlled by a microprocessor. Essential components are permanently controlled by the processor and thus faulty operation is prevented. After switching on, all instrument func- tions are checked during an automatic self-test routine.
  • Page 11 1 Introduction PHYSIOMED ELEKTROMEDIZIN AG • Conditions following radiothorium treatments, X-ray therapy • Conditions following laminectomy • Skin lesions (infections, inflammatory processes, naevi) • Tumescences at all stages (pre-/postoperative) • Blood coagulation diseases • In the area of the brain, spinal cord and eyes...
  • Page 12: Chapter 2 Controls And Indicators

    Chapter 2 Controls and Indicators The design of IONOSON-Basic allows for easy operation, in combination with a variety of functions. Be- cause of its small size, the instrument is very easy to transport. It has been designed for operation both in-...
  • Page 13: Data Selector <2

    Afterwards, the instrument will automatically reduce the intensity to zero. 2.3.1 Automatic Output Current Switch-off IONOSON-Basic features an automatic output current switch-off, activated in case the current flow of the electrodes is interrupted (electrode falls off, connector is disconnected from patient lead etc.). The mes- sage CHECK ELECTRODES I or CHECK ELECTRODES II will appear in the lower status bar and the cur- rent will be automatically turned down to a minimum basic current in the respective channel.
  • Page 14: Patient Current Indicator <5

    2 Controls and Indicators PHYSIOMED ELEKTROMEDIZIN AG 2.4 Patient Current Indicator <5> The diode at the left side of the symbol belongs to channel I, the diode at the right to channel II. The flashing of the indicators depends on the resistance in the two channels.
  • Page 15: Power Switch <8

    2 Controls and Indicators PHYSIOMED ELEKTROMEDIZIN AG 2.7 Power Switch <8> If you look at the instrument from the front, the Power Switch <8> is on its left side. By means of this switch, you can switch the instrument on and off.
  • Page 16: Manual Release Key Socket <11

    2 Controls and Indicators PHYSIOMED ELEKTROMEDIZIN AG Replacing the ultrasonic transducer during treatment will not affect the selected mode of treatment. The dose, however, will automatically be turned down to zero and has to be increased again when continuing the treatment.
  • Page 17: Chapter 3 Overview Of Current Modes

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG Chapter 3 Overview of Current Modes NOTICE All current modes of the instrument are emitted in the constant current mode unless other- wise specified (refer to Constant Current Operation on page 40). Only in simultaneous treat-...
  • Page 18: Simultaneous And Alternating Stimulation

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG • (FM) Frequency-Modulated Current on page 27 • (STOCH) Stochastic Current on page 29 • (T/R) Pulses with Adjustable Parameters on page 32 • (IFM) Modulated Interference Current on page 33 •...
  • Page 19 3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG Selection window for channel I With the Data Selector <2>, click first on the symbol R, set the pulse pause (1- 60 s) by turning and confirm by pressing. Now use the Data Selector <2> to click on the symbol T, set the pulse duration (1- 60 s or manual release key) by turning and confirm by pressing it.
  • Page 20 3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG • : Pressing the key triggers a single pulse of the preset increase duration T. Now use the Data Selector <2> to click on the symbol to return to the start menu.
  • Page 21: Low-Frequency Current Modes

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG Selection window for channel II First use the Data Selector <2> to click on the R symbol, turn to the left until the desired symbol (simultaneous or alternating stimulation) appears and press to confirm.
  • Page 22: Gmc) Galvanization With Microcurrent

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG Current Mode G: Direct Current Current Mode G: Medium-Frequency Interrupted Direct Current Parameters for Current Mode G Polarity switch to switch the polarity of the two electrodes; the sign in the intensity dis- play changes accordingly.
  • Page 23: Df) Diadynamic Current Diphasé Fixe

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG Parameters for Current Mode GMC Intensity display in microampere (1,000 µA = 1 mA) Polarity switch to switch the polarity of the two electrodes; the sign in the in- tensity display changes accordingly.
  • Page 24: Mf) Diadynamic Current Monophasé Fixe

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG Simulation fre- 100 Hz quency Pulse time 10 ms Release time 0 ms 3.3.4 (MF) Diadynamic Current Monophasé Fixe Definition Diadynamic current monophasé fixe acc. to Bernard Application • Pain treatment Menu for Current Mode MF...
  • Page 25: Cp) Diadynamic Current Modulé En Courtes Périodes

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG 3.3.5 (CP) Diadynamic Current Modulé en Courtes Périodes Definition Diadynamic current modulé en courtes périodes according to Bernard Application • Pain treatment • Resorption enhancement Menu for Current Mode CP Current Mode CP Parameters for Current Mode CP Polarity switch to switch the polarity of the two electrodes;...
  • Page 26: Ur) Ultra Stimulation Current

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG Menu for Current Mode LP Current Mode LP Parameters for Current Mode LP Polarity switch to switch the polarity of the two electrodes; the sign in the intensity dis- play changes accordingly.
  • Page 27: Hv) High-Voltage Current

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG Menu for Current Mode UR Current Mode UR: Monophase Current Mode UR: Biphase Parameters for Current Mode UR Polarity switch to switch the polarity of the two electrodes; the sign in the intensity dis- play changes accordingly.
  • Page 28 3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG Menu for Current Mode HV Current Mode HV: monophase Current Mode HV: biphase Parameters for Current Mode HV Polarity switch to switch the polarity of the two electrodes (the sign in the intensity dis- play changes accordingly);...
  • Page 29: Tens) Transcutaneous Electric Nerve Stimulation

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG Press the Data Selector <2>. The arrow symbol starts flashing. Press the Data Selector <2> and adjust the desired frequency (the desired frequency is displayed next to the f symbol) and press it for confirmation.
  • Page 30: Mens) Electric Nerve Stimulation With Microcurrent

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG How to Adjust the Pulse Time Click on the Ti symbol using the Data Selector <2>. The symbol starts flashing. Turn the Data Selector <2>, to adjust the desired pulse time and confirm by pressing it.
  • Page 31: Ig30) Pulse Galvanization 30

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG Current Mode MENS: monophase Current Mode MENS: biphase Parameters for Current Mode MENS Polarity switch to switch the polarity of the two electrodes; the sign in the intensity dis- play changes accordingly.
  • Page 32: Ig50) Pulse Galvanization 50

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG Menu for Current Mode IG30 Current Mode IG30: monophase Current Mode IG30: biphase Parameters for Current Mode IG30 Polarity switch to switch the polarity of the two electrodes ( the sign in the intensity dis- play changes accordingly);...
  • Page 33: Fm) Frequency-Modulated Current

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG Menu for Current Mode IG50 Current Mode IG50: monophase Current Mode IG50: biphase Parameters for Current Mode IG50 Polarity switch to switch the polarity of the two electrodes ( the sign in the intensity dis- play changes accordingly);...
  • Page 34 3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG Menu for Current Mode FM Current Mode FM: monophase Current Mode FM: biphase Parameters for Current Type FM Polarity switch to switch the polarity of the two electrodes ( the sign in the intensity dis- play changes accordingly);...
  • Page 35: Stoch) Stochastic Current

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG 3.3.14 (STOCH) Stochastic Current Definition Stimulation current with random pulse triggering Application • Circulatory stimulation • Analgesia with minimal stimulus adaptation Menu for Current Mode STOCH Current Mode STOCH: monophase Current Mode STOCH: biphase...
  • Page 36: Fas) Faradic Surge Current

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG NOTICE You can only change the pulse shape from triangle to square if the intensity in both channels is zero. Square pulses make the current mode more intense. 3.3.15 (FaS) Faradic Surge Current...
  • Page 37 3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG ately for both channels Pulse time Ti (1 .. 10 ms) and release time Ri (15 .. 150 ms); selection of the current mode, the basic parameters of the neofaradic current are set.
  • Page 38: T/R) Pulses With Adjustable Parameters

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG 3.3.16 (T/R) Pulses with Adjustable Parameters Definition Pulses with adjustable parameters Application • Treatment of peripheral paralysis • selective muscle stimulation Menu for Current Mode T/R Current Mode T/R: monophase Current Mode T/R: biphase...
  • Page 39: Medium Frequency Current Modes

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG 3.4 Medium Frequency Current Modes 3.4.1 (IFM) Modulated Interference Current Definition Medium frequency alternating current that features a outstanding compatibility and significant depth effect. Application • Pain therapy • Circulatory stimulation •...
  • Page 40: Amf) Amplitude-Modulated Medium-Frequency Current

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG NOTICE You can only shorten the release time if the intensity in both channels is zero. How to Program a Frequency Band Click on the symbol by means of the Data Selector <2>. The symbol starts flashing.
  • Page 41 3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG Current Mode AMF: Sinusoidal Waves Parameters for Current Mode AMF Frequency cycle (1 s / 30 s / 1 s; 1 s / 1 s / 1 s; 3 s / 3 s / 3 s; 6 s / 6 s / 6 s; 12 s / 12 s / 12 s;...
  • Page 42: Kots) Medium-Frequency Muscle Training

    3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG How to Adjust a Selective Frequency Click on the symbol by means of the Data Selector <2>. The symbol starts flashing. Turn the Data Selector <2> to the right until the SELECTIVE symbol is displayed.
  • Page 43 3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG Acoustic signal; makes pulses audible NOTICE Each time the intensity is incremented, the instrument puts out a pulse with a length of about 4 seconds. IONOSON-Basic...
  • Page 44: Chapter 4 Operation Of The Instrument

    PHYSIOMED ELEKTROMEDIZIN AG Chapter 4 Operation of the instrument The operating steps not directly relating to the therapy are described in the following paragraphs. 4.1 Mains and Battery Operation Two IONOSON-Basic models are available: • for mains operation only •...
  • Page 45: Economy Mode

    The Display <1> is no longer illuminated. Turning the Data Selector <2> will re-activate the illumination. 4.2 Start-up of the Instrument Before you can perform the first treatment with the IONOSON-Basic, you have to start the instrument up accordingly. How to Start up the Instrument Make sure that the instrument is switched off (the switch on the Mains Module <7>...
  • Page 46: Constant Current Operation

    Problem with ultrasound module during operation 4.5 Constant Current Operation The IONOSON-Basic can emit all current modes except G, GMC and MENS with constant current (CC) and with constant voltage (CV). The constant current operation (CC) is preset. The respective operation type is displayed in the lower status bar.
  • Page 47: Chapter 5 Therapy With Stimulation Current

    Only switch off the instrument AFTER you have removed the electrodes from the patient! 5.1 General Information With the IONOSON-Basic, you have the following options to access the stimulation current therapy: • direct access via the stimulation current menu (refer to...
  • Page 48: Modes Of Attaching The Electrodes

    5 Therapy with Stimulation Current PHYSIOMED ELEKTROMEDIZIN AG 5.2.1 Modes of Attaching the Electrodes The following techniques to attach electrodes are common in stimulation current therapy: Bipolar Here you apply two electrodes of equal size. This is the most common mode especially for pain therapy and circulatory stimulation.
  • Page 49: Preparations And Attaching The Electrodes

    5 Therapy with Stimulation Current PHYSIOMED ELEKTROMEDIZIN AG Type Diameter of the Electrode Effective Area Maximum Intensity for Currents with (cm) High Galvanic Component (mA) 5.4 Preparations and Attaching the Electrodes To prepare the electrodes, proceed as follows: Prior to attaching the electrodes, make sure that the intensity in both channels is turned down to zero.
  • Page 50: Therapy Using Low-Frequency Current Modes

    5 Therapy with Stimulation Current PHYSIOMED ELEKTROMEDIZIN AG Introduction to stimulation current therapy (ref. no. 00625). 5.5 Therapy Using Low-Frequency Current Modes 5.5.1 Monophase and Biphase Currents The majority of low-frequency current modes for pain therapy, circulatory stimulation, detonisation and mobilization can be applied in their basic monophase as well as biphase mode.
  • Page 51 5 Therapy with Stimulation Current PHYSIOMED ELEKTROMEDIZIN AG Current Mode G The current modes designated for iontophoresis are G and GMC. You find a detailed overview about the parameters adjustable under Parameters for Current Mode G on page 16 and...
  • Page 52: Pain Therapy, Hyperaemisation, Detonisation

    5 Therapy with Stimulation Current PHYSIOMED ELEKTROMEDIZIN AG Socket <11> on the front of the instrument and hand the manual release key to the patient. Slowly turn up the intensity of channel I with the Intensity Control Channel I <3>. The therapy time elapses.
  • Page 53: Muscle Stimulation Using Low-Frequency Currents

    5 Therapy with Stimulation Current PHYSIOMED ELEKTROMEDIZIN AG How to Perform a Treatment Switch on the instrument. Attach the electrodes to the respective stimulation points. From the Display <1>, select one of the above-mentioned current modes, e.g. STOCH. Select the parameters variable for the respective current mode (e.g. simulation frequency, biphase pulse output, pulse shape, etc.).
  • Page 54 5 Therapy with Stimulation Current PHYSIOMED ELEKTROMEDIZIN AG How to Perform a Treatment with Automatic Current Output Switch on the instrument. Attach the electrodes to the respective stimulation points. Position the patient to enable free move- ment of the stimulation area.
  • Page 55 5 Therapy with Stimulation Current PHYSIOMED ELEKTROMEDIZIN AG With the Data Selector <2> first click on the R symbol, rotate to the left until one of the manual control symbols appears and press to confirm. The symbols have the following meaning: •...
  • Page 56: Therapy With Medium Frequency Current Modes

    5 Therapy with Stimulation Current PHYSIOMED ELEKTROMEDIZIN AG <3> or Intensity Control Channel 2 <4>), until the muscle contraction is sufficient. The therapy time elapses. CAUTION Make sure you do not touch the electrodes when the current is turned up!
  • Page 57: Pain Therapy, Hyperaemisation, Detonisation With Amf Current

    5 Therapy with Stimulation Current PHYSIOMED ELEKTROMEDIZIN AG NOTICE This current mode requires both channel I and II at the same time. You cannot select an addi- tional current mode in channel II. How to Perform a Treatment with IFM Current Switch on the instrument.
  • Page 58: Muscle Stimulation With Medium Frequency Stimulation Currents

    5 Therapy with Stimulation Current PHYSIOMED ELEKTROMEDIZIN AG You find a detailed overview about the parameters adjustable under (AMF) Amplitude-Modulated Medi- um-Frequency Current on page 34. NOTICE This current mode can only be applied in channel I. You cannot set a second current mode in channel II.
  • Page 59 5 Therapy with Stimulation Current PHYSIOMED ELEKTROMEDIZIN AG How to Make a Treatment with Medium Frequency Stimulation Current KOTS Switch on the instrument. Attach the electrodes to the respective stimulation points. Position the patient to enable free move- ment of the stimulation area.
  • Page 60 5 Therapy with Stimulation Current PHYSIOMED ELEKTROMEDIZIN AG Select the modulation frequency. If necessary, alter the basic frequency F (10) Activate the acoustic signal if desired. (11) Set the therapy time. (12) Slowly turn up the intensity with the manual release key pressed (Intensity Control Channel 1 <3>...
  • Page 61: Chapter 6 Ultrasound Therapy

    6 Ultrasound Therapy PHYSIOMED ELEKTROMEDIZIN AG Chapter 6 Ultrasound Therapy IONOSON-Basic offers you three possibilities to start ultrasound therapy: • directly via the ultrasound menu • via the indications menu (refer to Indications Menu on page 59) • via an individual program (refer to...
  • Page 62 6 Ultrasound Therapy PHYSIOMED ELEKTROMEDIZIN AG Apply some ultrasound gel on the area of treatment. Choose SONO in the Display <1> . Select the duration of the treatment. Increase the dose to the desired value with the Data Selector <2>.
  • Page 63: Chapter 7 Simultaneous Therapy

    7 Simultaneous Therapy PHYSIOMED ELEKTROMEDIZIN AG Chapter 7 Simultaneous Therapy Simultaneous therapy means the treatment with ultrasound and stimulation current at the same time. The ultrasonic transducer is one pole of the stimulation current circuit, a plate electrode the other one.
  • Page 64 7 Simultaneous Therapy PHYSIOMED ELEKTROMEDIZIN AG Simultaneous Therapy You will find a description of the individual symbols in Ultrasound Therapy on page 55. Rotate the Data Selector <2> until you reach the SIM symbol. Now press the Data Selector <2>, rotate it one position further and press it once again to confirm.
  • Page 65: Chapter 8 Indications Menu

    8 Indications Menu PHYSIOMED ELEKTROMEDIZIN AG Chapter 8 Indications Menu Apart from direct access to the individual current modes, the indications menu offers you another possibil- ity to start the individual methods of treatment. With the indications menu, you may use proven stimula- tion current or ultrasound treatments.
  • Page 66 8 Indications Menu PHYSIOMED ELEKTROMEDIZIN AG NOTICE The page number is displayed on the right of the lower status bar. The left number iden- tifies the page selected, the right number the total number of pages. Indications Menu: Attaching the Electrodes Attach the electrodes to the respective stimulation points.
  • Page 67: Chapter 9 Individual Programs

    9 Individual Programs PHYSIOMED ELEKTROMEDIZIN AG Chapter 9 Individual Programs You do not have to set frequently applied forms of therapy every time you want to use them – you can simply save them as individual programs. Such a program is e.g. a certain current mode or an ultrasound therapy with all parameters including treatment time.
  • Page 68: Call Program

    9 Individual Programs PHYSIOMED ELEKTROMEDIZIN AG Turn the Data Selector <2> to the right and click on the symbol (save program). The symbol flashed and the cursor jumps to the first position of the selected program. Turn the Data Selector <2> to select the desired position and confirm by pressing it.
  • Page 69 9 Individual Programs PHYSIOMED ELEKTROMEDIZIN AG Turn the Data Selector <2> to the right and click on the symbol (delete program). The bin sym- bol flashes and the cursor jumps to the first position of the selected program. Turn the Data Selector <2> to select the desired position and confirm by pressing it.
  • Page 70: Appendix A Annex

    Maintenance measures on the instrument may be carried out only at the PHYSIOMED ELEKTROMEDIZIN AG authorised locations. If the work is carried out by such locations, a certification should be obtained from them about the type and extent of the work.
  • Page 71: Electromagnetic Compatibility

    If the vacuum application unit is connected and switched on, the current is generally released automatic- ally via the vacuum electrodes. As a note, a respective electrode symbol is displayed in the lower status bar. The following vacuum application devices can be operated together with IONOSON-Basic: • PHYSIOVAC-Basic Important By any means, refer to the Operating Instructions PHYSIOVAC-Basic (ref.
  • Page 72: Basic Settings

    Annex PHYSIOMED ELEKTROMEDIZIN AG A.7 Basic Settings In the Basic Settings menu, you can adjust the following instrument parameters: Symbol Meaning Contrast of the Display <1> Power threshold at which the instrument switches from continuous to pulsed power out- put in continuous mode...
  • Page 73: Technical Data

    Annex PHYSIOMED ELEKTROMEDIZIN AG A.8 Technical Data General Technical Data Protection class acc. to IEC 60601-1/ VDE 750 Teil1 I, type: BF CE characterization acc. to Council Directive concerning medical devices (93/42 EEC) Class acc. to Council Directive concerning medical...
  • Page 74 Annex PHYSIOMED ELEKTROMEDIZIN AG Ultrasound Frequency 1 MHz 3 MHz <= 6 <= 6 Bundle type Diverging Diverging Modulation Continuous ultrasound cont. Pulsed ultrasound: duty cycle (impulse/pause) 1:10; 1:5; 1:3; 2:5 Pulsed ultrasound: pulse frequency 100 Hz Adjustable switching between continuous and 0.5 W/cm²...
  • Page 75: Appendix B Scope Of Delivery And Accessories

    Alternatively you can send the electrodes to our service department for an inspection, which is liable to pay a service fee. B.1 Scope of Delivery (Mains-Operated Model) IONOSON-Basic as mains-operated model is supplied with the following accessories: Art.-No. Designation Number...
  • Page 76: Scope Of Delivery (Battery-Operated Model)

    Scope of Delivery and Accessories PHYSIOMED ELEKTROMEDIZIN AG B.2 Scope of Delivery (Battery-Operated Model) IONOSON-Basic as battery-operated model is supplied with the following accessories: Ref. No. Designation Quantity 02004 Combination therapy instrument IONOSON-Basic (battery-operated model) 00776 Mains cable 00344 Manual release key...
  • Page 77 Scope of Delivery and Accessories PHYSIOMED ELEKTROMEDIZIN AG Art.-No. Designation 00184 Filling pump 00413 Glove electrode (pair) 00156 Grip handle for electrodes with pushbutton 00155 Grip handle for electrodes 00136 Iontophoresis foil 18 x 12 cm; 1000 items 00776 Mains cable...
  • Page 78 Scope of Delivery and Accessories PHYSIOMED ELEKTROMEDIZIN AG Art.-No. Designation 00656 Plate electrode EF100 / 12 x 9 cm (red/blue) 00535 Plate electrode EF100 / 12 x 9 cm (red/red) 00565 Plate electrode EF100 / 12 x 9 cm (red/red) 1,5 m...
  • Page 79 Scope of Delivery and Accessories PHYSIOMED ELEKTROMEDIZIN AG Art.-No. Designation 00439 Viscose cover EF200 / 20 x 14 cm 00464 Viscose sponge for bow electrode Ø 2.0 cm 00463 Viscose sponge for bow electrode Ø 4.5 cm IONOSON-Basic...
  • Page 80: Appendix C Supplementary Documents

    Supplementary Documents PHYSIOMED ELEKTROMEDIZIN AG Appendix C Supplementary Documents C.1 Manufacturer’s Recommendations IONOSON-Basic...
  • Page 81: Index

    Index PHYSIOMED ELEKTROMEDIZIN AG Index MENS STOCH accessories TENS available alternating stimulation current modes ambient temperature low-frequency AMF (current mode) medium frequency application overview automatic switch-off current supply output current customer service basic settings batteries data selector handling delete battery operation...
  • Page 82 Index PHYSIOMED ELEKTROMEDIZIN AG GMC (current mode) using low-frequency currents with medium frequency currents high-voltage current HV (current mode) nerve stimulation 23 , 24 hyperaemisation notes general IFM (current mode) IFM current operation therapy output current IG30 (current mode) automatic switch-off...
  • Page 83 Index PHYSIOMED ELEKTROMEDIZIN AG stimulation alternating simultaneous stimulation current therapy medium frequency current modes therapy using low-frequency current modes stimulation currents medium frequency STOCH (current mode) stochastic current surge current faradic T/R (current mode) current mode technical data TENS (current mode)