Physiomed IONOSON-Basic Operating Instructions Manual page 37

Mains- / battery-operated
Table of Contents

Advertisement

PHYSIOMED ELEKTROMEDIZIN AG
NOTICE
Each time the intensity is incremented, the instrument puts out a pulse with a length of about
4 seconds.
How to Adjust the Pulse Time
(1)
Click on the Ti symbol using the Data Selector <2>. The symbol starts flashing.
(2)
Turn the Data Selector <2>, to adjust the desired pulse time and confirm by pressing it.
NOTICE
You can only extend the pulse time Ti if the intensity in both channels is zero. If you ex-
tend the pulse time, the current is experienced by the patient as more intense.
How to Adjust the Release Time
(1)
With the Data Selector <2>, click on the Ri symbol. The symbol starts flashing.
(2)
Turn the Data Selector <2>, to set the desired release time and press to confirm. The selected re-
lease time is displayed next to the Ri symbol.
NOTICE
Shortening the release time makes the current mode more intensive; extending it leads
to shiver-inducing frequencies.
You can only shorten the release time if the intensity in both channels is zero.
How to Change the Pulse Shape
(1)
With the Data Selector <2>, click on the symbol
(2)
Turn the Data Selector <2> to the right by one position to make the pulse shape change and press
it for confirmation.
NOTICE
You can only change the pulse shape from triangle to square if the intensity in both
channels is zero. Square pulses make the current mode more intense.
IONOSON-Basic
ately for both channels
Pulse time Ti (1 .. 10 ms) and release time Ri (15 .. 150 ms); selection of the current
mode, the basic parameters of the neofaradic current are set.
Pulse shape (triangle pulse, square pulse)
Acoustic signal; makes every pulse audible
3 Overview of Current Modes
. The symbol starts flashing.
31