PHYSIOMED ELEKTROMEDIZIN AG
(9)
With the Data Selector <2> first click on the R symbol, rotate to the left until one of the manual
control symbols appears and press to confirm. The symbols have the following meaning:
releasing the manual release key.
(10) Set the therapy time.
(11) Slowly turn up the intensity with the manual release key pressed (Intensity Control Channel 1
<3> or Intensity Control Channel 2 <4>), until the muscle contraction is sufficient. The therapy
time elapses.
CAUTION
Make sure you do not touch the electrodes when the current is turned up!
When using the manual release key, release a pulse each time you turn up the intensity!
(12) Release the button. If the manual release key is pressed again, the current will slowly rise automatic-
ally and will keep the set intensity as long as the manual release key is pressed. After releasing the
button, the intensity will automatically be reduced slowly.
(13) Signs of fatigue should make you stop the treatment (they may already occur after a few muscle con-
tractions)!
(14) After expiry of the timer, an acoustic signal will be issued. The intensity is turned down to zero auto-
matically. Remove the electrodes.
NOTICE
If the intensity (Intensity Control Channel I <3> or Intensity Control Channel II <4>)
is turned up during the time of release, a corresponding message will appear in the
lower status bar. Slowly turn down the intensity of the respective channel then. The
message disappears as soon as the previously set value is reached again.
The intensity can only be turned up effectively when the manual release key is de-
pressed.
How to Perform a T/R Treatment with the Manual Release Key
(1)
Switch on the instrument.
(2)
Plug the manual release key into the Manual Release Key Socket <11> at the front of the instru-
ment.
(3)
Attach the electrodes to the respective stimulation points. Position the patient to enable free move-
ment of the stimulation area.
(4)
Depending on the indication, use the Display <1> to select the current mode T/R.
(5)
If necessary, select the biphase pulse emission.
(6)
Activate the acoustic signal if desired.
(7)
If necessary, adjust the pulse time Ti.
(8)
Use the Data Selector <2> to click on the Ri symbol, turn it to the right until the manual release key
symbol
(9)
Set the therapy time.
(10) Slowly turn up the intensity with the manual release key pressed (Intensity Control Channel 1
IONOSON-Basic
: The duration of increase is determined by pressing the manual release key, the pause by
: Pressing the key triggers a single pulse of the preset increase duration T.
appears and press it to confirm.
5 Therapy with Stimulation Current