4 Operation of the instrument
In case of a monitoring note, first perform a self-test once or several times (refer to
page 39), and check if the problem still occurs afterwards.
Important
Inform your service partner if the monitoring note does not disappear even after several self
tests! Never perform a treatment when the proper function of the instrument is not assured!
List of Monitoring Notes
Monitoring Note
ET 101 – 106
ET 201 – 206
US 101 – 104
US 201 – 203

4.5 Constant Current Operation

The IONOSON-Basic can emit all current modes except G, GMC and MENS with constant current (CC) and
with constant voltage (CV). The constant current operation (CC) is preset. The respective operation type is
displayed in the lower status bar.
CAUTION
In CV operation, the safety shutdown of the electrodes (refer to
Switch-off
NOTICE
When switching off the instrument, the selection is set back from CV operation to CC opera-
tion.
All current modes can also be saved in CV operation.
How to Change from CC Operation to CV Operation
(1)
Switch on the instrument.
(2)
Choose the desired current mode on the start monitor (e.g. UR).
(3)
Click on the
output is changed to constant voltage (CV).
40
on page 7) is disabled!
symbol and turn the Data Selector <2> to the right by one position. The current
Meaning
Problem with stimulation current module during
selftest
Problem with stimulation current module during op-
eration
Problem with ultrasound module during selftest
Problem with ultrasound module during operation
Automatic Output Current
PHYSIOMED ELEKTROMEDIZIN AG
Function Check
on
IONOSON-Basic