Table 7-1 Abnormal Baseline Troubleshooting (Continued)
Symptom
Baseline noise, random
7
93
Troubleshooting
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Possible Cause
Corrective Action
Air in flow cell
Purge fluid path to remove
air.
Solvents not properly
Degas/sparge solvents
degassed or sparged
(see
Section 4.3, Solvent
Degassing).
Flow erratic, pump not
Prime the pump.
primed
Check for air in the pump,
failing seals.
Solvents contaminated
Use fresh solvent.
Column contaminated
Clean/replace column.
Dirty flow cell
Clean fluidic path (see
Section 6.1, Cleaning the
Fluidic
Analog output cable not
Properly connect cable.
properly connected
between 2410 and data
system or recorder
System improperly
Plug into different outlet on
grounded
different electrical circuit.
Use power conditioner.
Recorder voltage
Set recorder to correct
incorrect
voltage.
Radio frequency noise
Eliminate interference.
Path).