Symptom
Flat-topped peaks

7.2.4 Incorrect Qualitative/Quantitative Results

If a peak is incorrectly identified by a data system or integrator, make sure that the
retention time is correct.
If retention times are correct and peak resolution is good, the cause of qualitative and
quantitative errors is not likely to be chromatographic; it is more likely due to inadequate
sample preparation or faulty processing of the data (integration).
To troubleshoot problems with qualitation and/or quantitation, refer to
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Possible Cause
Corrective Action
Detector not zeroed
Auto zero detector
baseline.
Incorrect recorder input
Adjust recorder input
voltage
voltage, or adjust detector
output cable to correct
position.
Sensitivity too high
Select a lower sensitivity.
Scale factor too high
Select a lower scale factor.
(recorder only)
Sample concentration or
Decrease sample
injection volume exceeds
concentration or injection
voltage output of detector
volume.
Chromatography Troubleshooting
7
Table
7-4.
98