7.2.3 Poor Peak Resolution

Before you address problems with peak resolution, be certain that peaks elute at the
correct retention time. The most common causes of poor peak resolution can also appear
as retention time problems.
If peak retention times are correct, determine if poor resolution occurs:
• Throughout the chromatogram
• At a single peak pair
If efficiency of early peaks is poor, extra-column band broadening, such as autoinjector or
guard column failure, may be at fault. If peak efficiency is poor throughout the
chromatogram, post-column band-broadening or loss of column efficiency may be the
cause.
If only one peak in a chromatogram is badly-shaped, the peak component may be
interacting with the column through a chemical mechanism different from the components
in the other peaks. To troubleshoot this resolution problem, you need an understanding of
the separation chemistry.
Use
Table 7-3
to troubleshoot peak resolution problems that may be affecting your results.