7.4 Hardware Troubleshooting

This section describes symptoms, causes, and corrective actions related to the 2410
refractometer hardware. Use
lies within the instrument.
Table 7-6 Waters 2410 Hardware Troubleshooting
Symptom
7
Detector inoperative
Four-character display
LEDs fail to illuminate
Parameter LED
indicators do not light
Four-character display
shows odd characters or
FAIL
103
Troubleshooting
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Table 7-6
when you know the problem you have encountered
Possible Cause
No power at outlet
Check outlet by connecting
another electrical unit known
to be in working order and see
if it operates.
Fuse blown
Check that the fan and display
are operational; if neither,
replace fuse (see Section 6.2).
Weak electrical
Check connections.
connections
Fuse blown
Check/replace faulty fuse.
Faulty CPU board
Call Waters Technical Service.
Fuse blown
Check and replace faulty fuse.
Faulty Display board
Call Waters Technical Service.
Faulty CPU board
Call Waters Technical Service.
Faulty Power Supply
Call Waters Technical Service.
Corrective Action