3.3.5 Connecting Injection Trigger Signals

The 2410 refractometer accepts the following injection trigger signals from a manual
injector:
• Auto zero signal to automatically adjust the zero offset of the 2410 refractometer
each time the injector makes an injection
• Chart mark (inject start) signal from a contact closure signal with each injection
Each time the 2410 refractometer receives a signal from a manual injector, it performs the
corresponding auto zero or chart mark function.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Waters 2410 Refractometer
Connector B (Inputs and Outputs)
Red
White
OUT
+ –
CH
CH
1
2
Black
Making Non-IEEE-488 Signal Connections
1 Chart Mark +
1
2
2 Chart Mark –
3 Chassis Gnd
3
4 Polarity 1 +
4
5
5 Polarity 1 –
6 Chassis Gnd
6
7
7 Polarity 2 +
8
8 Polarity 2 –
9
9 Chassis Gnd
10
10 Int Out +
11
11 Int Out –
12
12 Chassis Gnd
3
55