Installation Site Requirements - Waters 2410 Operator's Manual

Differential refractometer
Hide thumbs Also See for 2410:
Table of Contents

Advertisement

Table 2-1 Installation Site Requirements
Parameter
Operating temperature range
Storage temperature range
Relative humidity
Storage humidity range
Bench space
Static electricity
Power
Surface orientation
Power Requirements
The 2410 refractometer, which operates over the range 100 Vac to 240 Vac, is shipped
from the factory with two 2.0 A fuses.
Caution:
To avoid electrical shock, power off the 2410 refractometer and unplug the
power cord from the rear panel receptacle before you replace a fuse.
Caution:
To reduce the risk of fire hazard, always replace the fuse with the same type and
rating.
The two fuses are located above the power input receptacle within the power input module
on the rear panel
(Figure
2-3).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Requirement
+15 °C to +32.2 °C (59 °F to 90 °F); avoid
direct exposure to sunlight and
heating/cooling vents.
–40 °C to 70 °C (–104 °F to 158 °F)
20% to 80%, noncondensing
0% to 90%, noncondensing
At least 11.2 in. (28.4 cm) wide × 24.8 in. (63
cm) deep × 8.2 in. (20.8 cm) high (includes 5
in. [12.7 cm] clearance at rear)
< 8 kV contact
Grounded ac, 100/240 Vac, 50/60 Hz
Level (ensures proper drip tray function)
Site Selection and Power Requirements
2
33