Abnormal Baseline Troubleshooting - Waters 2410 Operator's Manual

Differential refractometer
Hide thumbs Also See for 2410:
Table of Contents

Advertisement

Table 7-1 Abnormal Baseline Troubleshooting
Symptom
Baseline drift, rapid
7
Baseline drift, slow
Baseline drift, descending
91
Troubleshooting
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Possible Cause
Corrective Action
Column not equilibrated
Equilibrate column.
Detector not allowed to
Allow detector to warm up
warm up
until baseline is stable.
Warm-up time varies
based on sensitivity.
Solvent contaminated or
Use fresh solvent.
not HPLC grade
Solvent not properly
Degas the solvent. Cap the
degassed (rapid or slow
solvent reservoir.
drift)
Tubing contaminated
Clean the tubing using the
procedure in
Cleaning the Fluidic
Flow fluctuations (rapid
Fix pump problems,
or slow drift)
replace pump seals, check
valves.
Solvent contaminated
Use fresh, degassed
solvent.
Ambient temperature
Stabilize operating
fluctuations
environment temperature
enough to allow full
equilibration. Keep the
system away from air
conditioning vents, chance
breezes, and direct
sunlight.
Dirty flow cell
Clean flow cell
(Section 6.1, Cleaning the
Fluidic
Leaky flow cell
Call Waters Technical
Service.
Section 6.1,
Path.
Path).