Error Messages
Error Messages
-200
Execution error
An error associated with execution has been generated for which this
instrument cannot specify the error message. This code indicates that an error
associated with execution defined in 11.5.1.1.5, IEEE488.2 has occurred.
-123
Exponent too long
The magnitude of the exponent was larger than 32000.
-178
Expression data not allowed
Expression data is not allowed.
-170
Expression error
Expression is improper.
F
1070
Fan failed
Cooling fan hardware failure is detected. When this error occurs, contact
Keysight Technology's Sales and Service Office or the company from which you
bought the device.
21
Fixture circuit defective
A fault has occurred in a circuit to absorb back-emf of the 42842A/B. Confirm
the connection to the 42842A/B.
24
Fixture cover open
The protective cover of the 42842A/B is open. Close the cover.
23
Fixture OPEN det. defective
The opening and closing detection sensor of the 42842A/B protective cover
detects an error. When this error occurs, contact Keysight Technology's Sales
and Service Office or the company from which you bought the device.
22
Fixture over temperature
The temperature inside the 42842A/B bias test fixture exceeds the limit. Turn
off the DC bias and cool down the equipment.
30
Function type not available
The Ls-Rdc, Lp-Rdc and Vdc-Idc are not available when the 42841A is
connected.
G
-105
GET not allowed
GET is not allowed inside a program message.