Keysight E4980A Service Manual page 63

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

Troubleshooting
Function Specific Troubleshooting
Step 8. Use the softkeys to select the BIAS INTERFACE test.
Softkey
INCR +
DECR -
Step 9. Press the EXECUTE - TEST START softkeys to execute the BIAS INTERFACE
test.
Check the settings of S1 and S2 described in the
E4980A/E4980AL's LCD displays "Illegal Test Setup".
Step 10. Confirm the /RESET LED on the bias interface simulator turns ON.
Step 11. Confirm CS0 and CS1 LEDs on the simulator turns ON as the E4980A/
E4980AL's output signal is displayed on the LCD (Refer to
Step 12. Confirm that ADRS1-ADRS6 LEDs on the simulator turn ON in accordance
with the hexadecimal number displayed on the LCD. One of the 6 LEDs turns
ON in sequence as shown in
Step 13. Confirm that DO0-DO7 LEDs on the simulator turn ON in accordance with the
hexadecimal number displayed on the LCD. One of the 8 LEDs turns ON in
sequence as shown in
Figure 4-17
Bias Current Interface Function Test
Step 14. Confirm that "DI0 TO 7" and "DI8 TO 15" on the LCD of the E4980A/ E4980AL
display hexadecimal number "ff".
Keysight E4980A/E4980AL Precision LCR Meter
Function
Increments the test number of the Test Menu.
Decrements the test number of the Test Menu.
Figure
4-17.
Figure
4-17.
Step
1, if the
Figure
4-17).
63