Replacement Procedure
Power Inlet Assembly Replacement
Power Inlet Assembly Replacement
Tools Required
•
•
•
•
Removal Procedure
Refer to
Step 1. Remove the 3/4 Rack Module Top as described in
Replacement" on page
Step 2. Remove the PPMC PCA as described in
page
Step 3. Remove the A1 Mother Board as described in
Replacement" on page
Step 4. Disconnect the power cable connector (item 1) from the Power Supply
Assembly.
Step 5. Remove the TORX T20 screw (item 2) fastening the ground cable.
Step 6. Remove the two TORX T10 screws (item 3) fastening the Inlet Assembly.
Step 7. Remove the Inlet Assembly (item 4) from the Rear Panel.
Step 8. Detach the fuse box (item 5) from the original Inlet Assembly and remove the
fuse (item 6).
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
When you fasten the ground cable with the TORX T20 screw, set
screwdriver's torque to 21.1 in-lb.
When you fasten the Inlet Assembly with the two TORX T10 screws, set
screwdriver's torque to 8.7 in-lb.
Do not forget to insert the fuse in the new Inlet Assembly.
100
Torque screwdriver, TORX T20
Torque screwdriver, TORX T8
Torque screwdriver, TORX T10
Box torque wrench, 5/8 inch
Figure 6-14
for this procedure.
73.
84.
86.
"3/4 Rack Module Top
"PPMC PCA Replacement" on
"A1 Mother Board