microspeed® uni motor system
Uncoupling a blade from the shaver handpiece
A
Fig. 21
Pull back unlocking sleeve E at the handpiece in the direction of the
arrow; remove blade A, holding it at cap B.
Connecting a suction tube to shaver handpiece GD682
36
37
Fig. 22
Legend
A Suction tube
36 Suction control lever
37 Shaver handpiece
38 Tube olive
To connect suction tube A to shaver handpiece 37: Push suction
tube A (GF572 or GF573) of the surgical suction device over tube
olive 38.
22
E
B
38

4.2 Function checks

Prior to each use, inspect the entire unit for functionality and proper
condition.
Verify that all components to be used function properly and are
properly connected.
The connected motor type must be indicated in the appropriate motor
control field on the display.
Set the motor release switch to its "ON" position, if necessary.
The motor to be used is released.
Actuate the foot or hand control down to the limit stop.
The motor starts and reaches the maximum speed indicated in the
motor control field in display 2.
The motor audibly runs smoothly and at a constant speed.
The qualitative display of the actual motor speed in the motor control
field is fully illuminated.
Enable coolant pump 4 in the respective motor control field or via the
function key of the foot or hand control device.
A
Coolant pump 4 is started by activating the motor.
Once the motor was active for more than two seconds, a special drive
control prevents further coolant from dripping out.