5.1.2 Manual cleaning/disinfecting
Apply cleaning and disinfectant solution.
Wipe away cleaning and disinfectant solution with a sponge or cloth
that has been moistened in clean water.
Inspect surfaces, cavities, lumens and openings for visible debris. If
necessary, repeat the cleaning/disinfection process.
Dry the product with a lint-free cloth.
5.1.3 Control, care and inspection
Inspect the product after each cleaning and disinfecting cycle to be
sure it is clean, functional and undamaged.
5.1.4 Storage
Store reprocessed products under conditions as germ-free as possible,
in a dry, dark, cool and dust-protected room.
5.2 Processing the microspeed® uni motors GD674,
GD676, GD677, GD678, GD679 and motor cables
GD672 and GD673
5.2.1 Preparations
Prior to first mechanical cleaning/disinfection: Mount ECCOS® storage
device GB578R (for GD674, GD676, GD677, GD678, GD679) and
GB689R (for GD672, GD673) in a suitable basket tray (e.g. JF214R).
Mount the storage devices in such a way that the cables can be coiled
in a circle with a bending radius of not less than 10 cm. Do not kink
the cables.
Separate the motor from the motor cable prior to reprocessing.
Carry out non-fixating/NaCl-free pre-cleaning immediately after use.
Remove all attached components such as adapters, tubes, etc.
Reprocess the product immediately after use.
Do not clean the product by ultrasound treatment under any
circumstances.
Put away the product in dry condition.
5.2.2 Cleaning/Disinfection
Damage to the product due to inappropriate cleaning/
disinfecting agents and/or excessive temperatures!
Only use cleaning and disinfecting agents that do not
CAUTION
affect softeners (e.g. silicone), according to the
manufacturers' instructions for use.
Do not exceed the maximum allowable cleaning
temperature of 55 °C.
Use cleaning/disinfecting agents that are suitable for the product.
Always
follow
the
manufacturer's
concentration, temperature and exposure time.
Avoid encrustation of residues/proteins (e.g. caused by aldehyde/
alcohol).
Only use bactericidal, fungicidal and virucidal disinfecting agents.
Preferably use thermal disinfecting processes.
instructions
regarding
39