B. Braun microspeed uni Instructions For Use Manual page 10

Aesculap power systems
Table of Contents

Advertisement

Available languages

Available languages

Aesculap Power Systems
microspeed® uni motor system
microspeed® uni motor cable for foot control GD672
Fig. 4
Legend
30 Motor release switch
Note
This motor cable can only be used in connection with a foot control.
The motor coupled to this motor cable can only be activated if motor
release switch 30 is in the "ON" position. When two motor cables GD672
are used at the same time, a motor can be activated only if it has been
released (motor release switch 30 in position "ON") while the other motor
is deactivated (motor release switch 30 in position "OFF").
8
30
microspeed® uni motor cable with hand control GD673
29
Fig. 5
Legend
25 Motor release switch:
Position "ON": Motor released for operation
Position "OFF": Motor locked against operation
26 Hand control function button:
Short actuation: Switching on/off the coolant pump
Pressing for more than three seconds (with the motor inactive):
Change to Device settings menu, siehe Kapitel Changing the settings
in the Device settings menu through hand control
29 Lever:
Actuator for motor speed setting
Note
This motor cable can not be used in connection with a foot control.
The motor coupled to this motor cable can only be activated if the motor
release switch is in its "ON" position.
25
26