NOTICE
Contact of the battery charger with magnetic data storage devices
Wiping of the data storage device.
► Do not place the battery charger on credit cards, diskettes, audio or video cassettes.

4.6 Information on Alignment/Adjustment

CAUTION
Use of unsuitable prosthesis components
Falling due to unexpected behaviour of the product or breakage of load-bearing components.
► Use the product only in combination with components listed in the section "Combination possibilities" (see
Page 8).
CAUTION
Improper assembly of the screw connections
Falling due to breakage or loosening of the screw connections.
► Clean the threads before every installation.
► Apply the specified tightening torque values for installation (see the section "Technical data" see Page 34).
► Observe the instructions for securing the screw connections and the use of the correct length.
CAUTION
Incorrectly secured screws
Falling due to breakage of load-bearing components caused by screw connections coming loose.
► After completing all settings, the set screws in the tube adapter must be secured before they are tightened to
the specified torque.
► The screws in the clamp bracket must not be secured but only tightened to the specified torque.
CAUTION
Incorrect alignment or assembly
Falling due to damage to the prosthesis components.
► Observe the alignment and assembly instructions.
CAUTION
Errors during prosthesis alignment
Falling due to breakage of load-bearing components.
► At maximum flexion (reached under full load!), it is essential to maintain a minimum distance of 3 mm (1/8")
between the hydraulic unit and the socket.
► At maximum flexion and insofar as contact with the frame of the knee joint cannot be avoided (in case of volu­
minous residual limbs), the socket must lie flat against the frame. Soft cushioning on the socket will assist in
keeping the socket flat.
CAUTION
Insufficient insertion depth of the tube adapter
Falling due to breakage of load-bearing components.
► Insert the tube adapter at least 40mm to ensure operational safety.
► The patient must be seated for length adjustments.
Kenevo 3C60/3C60=ST