Error Messages - Chroma 19020 User Manual

Multi-channel sync hipot tester
Hide thumbs Also See for 19020:
Table of Contents

Advertisement

Description:

6.6 Error Messages

The error messages stored in error queue will be returned in the way of first in first out
(FIFO) which means the first error message returned is the first one being saved.
When error messages exceed 30, the last one be stored in the error queue will be –350
"Queue overflow". It means the error queue is unable to store another error message
until the error messages are extracted.
If no error is generated, +0 "No error" will be stored in the first position of error queue.
-101
Invalid character
There is invalid character in the command.
-102
Syntax error
The syntax is error due to wrong character is used when piping the command.
-103
Invalid separator
There is invalid separator in the command string.
-104
Data type error
The parameter format is incorrect.
-108
Parameter not allowed
The device receives unallowable parameter.
-109
Missing parameter
The parameter is missing.
-111
Header separator error
The command header separator is incorrect.
-112
Program mnemonic too long
The simple command program header exceeds 12 characters.
-113
Undefined header
The device receives undefined command header.
-114
Header suffix out of range
The command header suffix variable is out of range.
-120
Numeric data error
The numeric parameter is incorrect.
-141
Invalid character data
The device receives invalid character data.
-151
Invalid string data
The device receives invalid string data usually the missing double quote.
-158
String data not allowed
The device receives unallowable string parameter.
-168
Block data not allowed
The device receives unallowable block parameter.
-171
Invalid expression error
The device receives invalid parameter of math expression.
-178
Expression data not allowed
The device receives unallowable parameter of math expression.
-200
Execution error
Error occurs when executing a command.
-203
Command protected
The device does not accept the command.
-221
Settings conflict
The tester returns 1990.0.
The returned 1990.0 means the SCPI version supported by this
device is 1990.0.
GPIB/RS232 Interface (IEEE-488.2)
6-37