Setting Fail Lock - Chroma 19020 User Manual

Multi-channel sync hipot tester
Hide thumbs Also See for 19020:
Table of Contents

Advertisement

Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
User's Manual
Enter the PASSWORD and press ENTER, the LOCK box returns to normal indicating the
KEY LOCK is cancelled.

4.4.5 Setting FAIL LOCK

The way to set FAIL LOCK:
In SYSTEM screen, move the highlight to FAIL LOCK and press ENTER to go to FAIL LOCK
setting screen as shown below:
FAIL LOCK
FAIL KEY:
USER PASSWORD:■
1.
Enter the PASSWORD when in FAIL LOCK screen.
2.
When FAIL LOCK is ON, the LOCK text is reversed to indicate the host is in parameter
FAIL LOCK mode. The "RECALL", "OFFSET", "GET Cs" in [TEST] and "PROGRAM",
"STORE/RECALL" in [MAIN INDEX] as well as the "TEST CONTROL", "SYSTEM
CONFIG", "KEY LOCK", "CHANGE PASSWORD" and "CALIBRATION" in [SYSTEM]
are all invalid for setting.
3.
When FAIL LOCK is set and the test result is FAIL, all keys are invalid except the
Function Key [CLEAR] and STOP in TEST screen. It is necessary to press the Function
Key [CLEAR] to enter the FAIL LOCK password to continue the test.
The way to release FAIL LOCK:
If the LOCK box is reversed in power on screen, the FAIL LOCK function can be released.
Press SYSTEM to select FAIL LOCK as shown below:
4-12
EXIT
REMOTE LOCK
CORR
ERROR