7.1 Entering Calibration Screen
When in SYSTEM screen, move the highlight to CALIBRATION and press ENTER to go to
CALIBRATION setting screen as shown below:
CHANGE PASSWORD
CAL. PASSWORD︰■
Key-in the calibration password 7 9 3 1 to enter into the calibration mode.
7.2 Selecting the Channel for Calibration
CALIBRATION
A/D REFERENCE
CHANNEL: 1
VOLTAGE
1.500V
As the figure shown above, use Function Key [NEXT CH] and [PREV CH] to select the
channel to be calibrated and the screen will show the channel under calibration.
REMOTE LOCK
CURRENT
1.500V
REMOTE LOCK
Calibration Procedure
EXIT
CORR
ERROR
NEXT CH
PREV CH
CORR
ERROR
7-3