ACV 5kV 60Hz FULL (4kV) Calibration Screen:
CALIBRATION
ACV 5kV 60Hz FULL (4kV)
CHANNEL: 1
OUTPUT
4.000kV
STANDARD
3.998kV
DEF=4kV
1.
Press START to output voltage and read the data from the high voltage meter.
2.
Press numeric key 0
3.
Press ENTER to confirm the input.
4.
Press STOP to stop high voltage output.
5.
Press ▼ to go to DCV 6kV AGC OFFSET (0.1kV) calibration.
7.4.2

DCV Calibration

Connect a DCV high voltage meter to the Tester with high voltage terminal
connected to HV OUTPUT and low voltage terminal connected to RTN/LOW of the
channel to be calibrated.
DCV 6kV AGC OFFSET(0.1kV) Calibration Screen:
CALIBRATION
DCV 6kV AGC OFFSET (0.1kV)
CHANNEL: 1
OUTPUT
0.100kV
STANDARD
0.101kV
DEF=0.1kV
1.
Press START to output voltage and read the data from the high voltage meter.
2.
Press numeric key 0
3.
Press ENTER to confirm the input.
4.
Press STOP to stop high voltage output.
5.
Press ▼ to go to DCV 6kV AGC FULL (4kV) calibration.
READING
4.012kV
REMOTE LOCK
. ~ 9 to input the reading of high voltage meter.
READING
0.102kV
REMOTE LOCK
. ~ 9 to input the reading of high voltage meter.
Calibration Procedure
NEXT CH
PREV CH
CORR
ERROR
NEXT CH
PREV CH
CORR
ERROR
7-7