Routine Maintenance
12.
Press [Utility] to access the utility functions, after initialization.
O O
13.
Press [
] until Lamp Alignment appears on the display.
14.
Press [Align] to display an energy graph as shown below.
15.
Use a slotted screwdriver to adjust the alignment screw [Item #5] until
the graph displays its maximum value.
16.
Press [Esc] twice to return to the standard display.
17.
If desired, reset the hours of lamp operation to zero (refer to the
GENESYS 20 Operator's Manual).
18.
Replace the door of the lamp compartment and allow the lamp to warm
up for at least 30 minutes.
5 - 3

Esc