9.3.5.2

Spring

3.
Unscrew the lamp holding knob (turn counterclockwise).
4.
Remove the socket from the assembly.
REPLACEMENT
5.
Position the socket on the entrance slit/illumination assembly mounting
screw. Make sure the screw is properly positioned and not bent or
squeezed.
6.
Screw the lamp holding knob onto the mounting screw (turn clockwise).
7.
Connect the lamp cable to the main circuit board.
8.
Replace the main cover assembly (refer to "Section 9.1.3 - Cover
Assembly").
REMOVAL
1.
Remove the main cover assembly (refer to "Section 9.1.3 - Cover
Assembly").
2.
Disconnect the lamp cable from the main circuit board.
3.
Remove the lamp socket (refer to "Section 9.3.5.1 - Lamp Socket").
4.
Pull the spring out of its mounting hole. This may require some force.
5.
Clean out as much of the old RTV adhesive as possible.
REPLACEMENT
6.
Apply RTV adhesive to the end of the spring and press it into its
mounting hole.
7.
Replace the lamp socket (refer to "Section 9.3.5.1 - Lamp Socket").
8.
Connect the lamp cable to the main circuit board.
9.
Replace the main cover assembly (refer to "Section 9.1.3 - Cover
Assembly").
9 - 19
Dissassembly and Replacement