Disassembly and Replacement
4.
Reach under the front of the unit and press the two clips toward the
sides of the unit and up. Figure 9.3 shows a bottom view of the base
assembly indicating the location of the two clips.
Figure 9.3 - Base Assembly: Bottom View
WHEN LIFTING THE COVER ASSEMBLY, TAKE CARE NOT TO
PULL ON THE WIRES CONNECTING THE KEYPAD AND DISPLAY
TO THE MAIN CIRCUIT BOARD. FAILURE TO OBSERVE THIS
CAUTION MAY RESULT IN DAMAGE TO THE WIRING HARNESS.
5.
Carefully lift the front of the cover assembly so the upper rear portion of
the cover moves away from the connectors on the edge of the main
circuit board.
6.
Carefully lift the cover assembly to free the back of the assembly from
the two screws located at the back of the base assembly.
7.
Continue to lift the cover to free the back panel which fits in a track in
the cover assembly.
9 - 4
CAUTION