2.
Turn the instrument over and lay it on its top.
Figure 9.12 - Genesys 20: Bottom View
3.
Remove the Sector gear and Limit Switch Cover.
4.
Remove the two screws which secure the limit switch.
5.
Remove the two wires attached to the limit switch terminals.
REPLACEMENT
6.
Attach the two wires to the limit switch.
7.
Locate the limit switch in position. Make sure the wires are routed
under the switch and spring. The tie wrap on the wires must be trapped
under the locking notch as indicated in Figure 9.12 after the wire is fed
through the hole in the base assembly.
9 - 23
Dissassembly and Replacement