Sample Compartment Cover Hinge Mounting - Thermo Scientific GENESYS 20 Service Manual

Hide thumbs Also See for GENESYS 20:
Table of Contents

Advertisement

2.
Remove the sample compartment cover from the sample compartment
hinge.
a.
Using a very small screwdriver, gently pry up on the center of
each hinge tab while gently pulling the cover slightly away
from the hinge.
b.
Slide the cover off the hinge tabs.
3.
Remove the hinge assembly from the main cover assembly.
Figure 9.6 - Sample Compartment Cover Hinge Mounting
a.
One at a time, press each hinge pin out of its mounting hole so
it catches on the outside of the plastic and does not return to
the hole.
b.
Slide the hinge away from the main cover assembly.
REPLACEMENT
4.
Carefully slide the hinge into position on the cover assembly until the
hinge pins lock into their locating holes.
5.
Carefully slide the sample compartment cover onto the hinge until the
tabs snap into position.
6.
Replace the main cover assembly (refer to "Section 9.1.3 - Cover
Assembly").
9 - 7
Dissassembly and Replacement