INSTALLATION

4
INSTALLATION
Before installing the analyzer, read through this manual carefully and take note of all
warnings.
30
Do not open the casing while powered-up in a hazardous
Special care must be taken when the analyzer is
The XTC601-EX versions must only be installed by
suitably qualifi ed personnel and in accordance with the
instructions provided and the terms of the applicable
product certifi cates.
Maintenance and servicing of the product must only be
carried out by suitably trained personnel or returned to an
approved Michell Instruments' Service Center.
Before the cover is refi tted, the fl amepath/threaded joint
between the cover and body must be thoroughly wiped
clean of dirt, grit or other foreign substances and then a
thin coating of an approved form of non-setting grease
applied to joint/threads. Ensure gasket is free from
Threaded covers must be screwed down until fully tight
(at least 7 full turns) and then locked in position with the
XTC601 User's Manual
area.
energized.
damage.
hex key provided.
97400 Issue 3.1, January 2017