Hazardous Area Safety

Appendix F of this manual refers to the EX version and the Hazardous Area Certification of this
product.
This product is fitted with a marking label that contains Hazardous Area information pertinent to the
suitable location and installation.
During all installation and operation activities local regulations and permitted working routines must
be observed. Installation should only be performed by competent personnel and in accordance with
IEC/EN 60079-14:2008 or local equivalent.
Repair and servicing of this equipment must only be carried out by the manufacturer.
Michell Instruments
This product is certifi ed safe for use in a Zone 1 and Zone 2 area
only. This product must not be installed or used within a Zone 0
This product must not be operated within an explosive
atmosphere greater than 1.1 bar absolute.
This product must not be operated with ambient oxygen levels
above 21% in a hazardous area.
This product must not be operated outside of
the temperature range of -20 to +55°C (-4 to +131°F).
WARNING:
area.
WARNING:
WARNING:
WARNING: