XTC601 User's Manual
2.7.1
Settings Page
The analyzer is microprocessor-based and, as such, has settings and features accessible
to the user.
Select the parameter required. The options will be highlighted and can be toggled
between by pressing the
which, if only one analyzer is connected to your system, should be set to '
Parameter
FIELD CAL
FIELD CAL 2
(HCG version only
FIELD CAL 3
(HCG version only
EXT COMP
(Standard version
)
only
LIMIT
0-100%
Modbus ID
NOTE: The HCG settings page does not include
Michell Instruments
FIELD CAL
EXT COMP
LIMIT 0-100%
MODBUS ID
Settings page (Standard version only)
Figure 19
button. These are all
ENTER
Description/Operation
When enabled, the field calibration corrections will be
applied. when disabled, the factory calibration will be
used. For HCG version, this option only applies to Phase 1.
It is automatically turned off when Michell or field
calibrations are being performed
When enabled, the field calibration corrections for phase
2 will be applied. When disabled, the factory calibration
)
will be used.
When enabled, the field calibration corrections for phase
3 will be applied. When disabled, the factory calibration
)
will be used.
Turn externals sensor compensation on or off
It is automatically turned off when Michell or field
calibrations are being performed
Limits the % Target Gas so that any drift below 0.00 and
above 100.00 (for suppressed zero) is not visible
mA outputs are also limited accordingly
Unit's network address for Modbus communications
Table 5
ON/OFF
ON/OFF
ON/OFF
1/128
ON/OFF
Settings parameters
EXT COMP
OPERATION
except for Modbus ID
'.
1
Options
ON/OFF
ON/OFF
ON/OFF
ON/OFF
ON/OFF
1-128
option.
17