Therapy Chart; Powered Cpr - Arjohuntleigh RotoProne User Manual

Table of Contents

Advertisement

Therapy Chart

Use the Therapy Chart to view cumulative therapy time in a bar graph format.
1. Press the Therapy Chart button. The Therapy Chart
screen will appear. Therapy time will be displayed in a
bar graph showing percentages of total therapy time in
prone and supine position.
2. Press the up and down arrow buttons to search for a
specific date.
3. Press the Done button to return to the Therapy
Meters screen.

Powered CPR

Use the CPR button on the screen or frame to automatically move the patient surface to level supine position
within 40 seconds.
Take care to ensure all lines and tubes remain attached when powered CPR is activated. As with
any medical device, familiarity with the RotoProne Unit can save time in an emergency situation.
Caregivers and other hospital personnel are encouraged to become familiar with the CPR function
and the procedure required to access the patient in case of an emergency.
1. Call for assistance. Close any open hatches. Ensure hatch latches are secure.
2. Press and hold either:
The CPR button on the main display (available in any screen)
or
The red and white CPR button on the front of the
unit. The CPR Mode screen will appear.