Arjohuntleigh RotoProne User Manual page 36

Table of Contents

Advertisement

ALARM CONDITION
Obstruction - Patient Surface Raising /
Lowering
Buckle / Hatch / Hoop Open Alarm
Partial menu shown. Use the scroll bar to scroll
down to see all potential alarm conditions.
WHAT IT MEANS
There is a possible
Follow the instructions on
obstruction in the
the screen. Use the manual
way of the patient
switches located behind the
surface.
small door at the foot of the
unit to raise the patient surface
off the obstruction. Check for
and remove any obstructions.
Press Cancel button to resume
patient surface raising /
lowering.
Obstruction error can occur
if someone or something is
leaning or standing on the
frame.
One or more of the
Follow the instructions on
following is open or
the screen. Check each
not secure:
item listed. Once all pack,
Proning Pack
hatch or hoop conditions are
Buckle(s) Upper Body
corrected, therapy will resume
/ Midline / Lower
automatically.
Body Hatch
Head Hoop
If Proning Pack Buckle is not
The screen will detail
secured and patient is in the
which item(s) are
prone position, the side support
open or loose and
packs may need tightening. If
need correction.
the patient can tolerate, return
the surface to 0° supine using
the CPR button on the frame.
Then, use the crank to tighten
side support packs, tighten
the buckles and resume prone
therapy.
2-24
CORRECTIVE ACTION