Cleaning the Citadel Patient Therapy System While in Use

1. If possible, remove the patient from the bed prior to cleaning. Daily
care and cleaning consists of wiping down all surfaces and side rails
(as needed) during patient bathing.
2. Follow care and cleaning instructions for the specific patient support
surface in use.
3. Unplug the Citadel Patient Therapy System from wall outlet.
4. Inspect the power cord for any signs of wear or damage. The Citadel
Patient Therapy System should not be operated with a worn or
damaged power cord. Contact ArjoHuntleigh if damage is found.
5. Using a cloth dipped in warm soapy water or approved hospital
disinfectant (diluted according to manufacturer's instructions), wipe
the surfaces of the Citadel Patient Therapy System. Rinse with plain
water.
Do not allow fluid to penetrate the Citadel Patient Therapy
System control panels.
6. Allow all components to dry completely before returning to use.
7. Inspect all parts of the Citadel Patient Therapy System for damage
prior to returning the unit to use. Contact ArjoHuntleigh for service or
replacement.
8. Plug bed into wall outlet and adjust settings.
9. With bed plugged into the wall outlet, press and hold the On / Off
button for two seconds to turn on the control unit.
Cleaning and Maintenance Between Patients
1. Unplug the Citadel Bed Frame System from wall outlet. Using a cloth
dipped in warm soapy water or approved hospital disinfectant (diluted
according to manufacturer's instructions), wipe the surfaces of the
Citadel Patient Therapy System. Rinse with plain water.
2. Allow all components to dry completely before returning to use.
Do not allow fluid to penetrate the Citadel Patient Therapy
System control panels.
3. Inspect all parts of the Citadel Patient Therapy System for damage
prior to returning to use. Contact ArjoHuntleigh for service or
replacement.
When not in service, bed frame must remain plugged in to
the wall outlet to maintain battery charge.
30
Citadel Patient Therapy System