PoWeR-DoWn PRoCeDuRes

The RotoProne Therapy System is not recommended for patient transport. Should there be an emergency situation where
hospital personnel transport the patient while on the RotoProne Unit, care should be taken in doorways and elevators or in
any situation where casters can get caught. A minimum of two persons are recommended for emergency transport.
Care should be taken prior to RotoProne placement to ensure the unit fits in facility elevators. Use the following
procedures to power down the RotoProne for intra-facility transport (for storage or maintenance) or for shipping
between facilities.

Power-Down for Intra-Facility Transport / Storage

1. Discontinue therapy.
2. Ensure patient surface is at 0° supine.
3. Push Lock Pin in.
4. Transfer patient from unit.
See Patient Care section of this manual for detailed instructions on patient transfer.
5. Press and hold OFF button on Home screen.
Shutdown Information screen will appear and alarm
will sound.
Alarm will sound until one of the buttons on the
screen is pressed.
All settings will be held in memory and restored
when the unit is plugged in.
6. Press Transport button (on the left side of the screen).
If the patient surface is not already leveled and
lowered, the Lock Pin Out screen will appear.
If the patient surface is leveled and lowered and the
Lock Pin is already pushed in, the Transport Height
Information screen (described as follows in Step 9) will
appear.
7. Pull Lock Pin. Rotate and Lower Information screen
will appear.
5-1