Cleaning and Disinfection Instructions ........24 Battery Instructions ..............28 Care and Preventative Maintenance ........29 Troubleshooting ..............33 Technical Specifications ............34 Dimensions ................36 Labels ..................38 Approvals and List of Standards ..........40 Electromagnetic Compatibility ..........40 Parts and Accessories ............42 ArjoHuntleigh Addresses ..........Last page...
Thank you for purchasing ArjoHuntleigh equipment. A service routine must be performed on your Miranti Your Miranti lift bath trolley is a part of a series of lift bath trolley every year by qualified personnel to quality products designed especially for hospitals, make sure the safety and operating procedures of your nursing homes and other health care use.
NOTE WARNING All materials used in the Miranti lift bath trolley complies with established biocompatibility To prevent cross-contamination, always standards and have been tested accordingly follow the disinfection instructions in this (ISO 10993-5).
4 Read the IFU. 4 Make sure that the battery is fully charged. 5 Disinfect the product according to section Miranti Lift Bath Trolley Directions Cleaning and Disinfection Instructions on page 24. The driving direction during the transfer is according to the illustration below.
Fig. 2 Hand Control The hand control has an integrated hook which makes it easy to attach to various parts of the Miranti lift bath trolley. (See Fig. 2) • Unhook the hand control and hold it in your hand.
Side Guard and Patient Security Grip Fig. 1 (3 steps) The side guard, which prevents the patient falling to the side and the patient security grip, which the patient can hold, are integrated in a hinged bar. 1 Lift the bar upwards. (See Fig. 1) Fig.
(See Fig. 6) Emergency Stop Fig. 7 If, for any reason, the Miranti lift bath trolley does not respond to the control buttons, you can quickly stop lift movement by pressing the button on the battery and disconnect it. (See Fig. 7) Contact your local ArjoHuntleigh representative before re-using the Miranti lift bath trolley.
Emergency High/Low Fig. 8 (2 steps) If, for any reason, the Miranti lift bath trolley does not respond to the control buttons, raise or lower the Miranti lift bath trolley by using the Emergency High/Low. 1 Unplug the hand control. (See Fig. 8)
The Miranti lift bath trolley has two safety belts, one to be used over the chest and one over the legs. They are identical and are used in the same way.
Page 16
4 Buckle the safety belt by inserting the buckles into each Fig. 5 other. Make sure the fastening clip has the arrow shaped side upwards (A). (See Fig. 5) 5 Unbuckle the safety belt by pressing the spring knobs (A), Fig.
Page 17
When not in use (1 step) Fig. 12 1 Make sure that the safety belts are buckled over the stretcher when the Miranti lift bath trolley is not in use. (See Fig. 12). Remove (1 step) Fig. 13 1 Pull the hook rope (upper safety belt only) and the safety...
3 Undress the patient in bed and dress him/her in a bathrobe, preferably one with buttons up the back. 4 Choose one end of the Miranti lift bath trolley to function as the backrest and fasten the hook rope in that end. If possible, approach the patient from the right side of the Miranti lift bath trolley for transfer to the bath.
Page 19
Let the patient put his/her hands on the patient security grip. (See Fig. 4) 13 Raise the Miranti lift bath trolley’s stretcher enough to move it away from the bed. 14 Release the Miranti lift bath trolley central brake.
Page 20
Fig. 1 2 Fold down the bed rail. 3 If the Miranti lift bath trolley comes from a bath: place a towel on the bed to avoid making the bed wet. 4 Adjust the Miranti lift bath trolley height so the stretcher is at a safe distance from the bed.
Page 21
12 Raise stretcher to separate it from the bed. 13 Release the central brake. 14 Pull the Miranti lift bath trolley away from the bed. 15 Turn the patient so he/she lays on the back. 16 Fold up the bed rail.
Bath the Patient Read the separate instructions for your bathing Fig. 1 equipment. Power Drive (9 steps) 1 Raise the stretcher to clear the edge of the bath. (See Fig 1) Fig. 2 2 Activate the power drive and move the patient along the side of the bathtub and finally push transversely to it.
Page 23
8 Dry the patient, as well as the underside of the stretcher, to Fig. 5 prevent water from dripping onto the floor prior to moving the Miranti lift bath trolley away from the bath. Raise the stretcher. (See Fig 5) Fig.
Cleaning and Disinfection Instructions For the best results, only use ArjoHuntleigh WARNING branded disinfectant. If you have any questions regarding disinfecting the equipment or wish to order disinfectant (see section Parts and Accessories on page 42), contact you local ArjoHuntleigh representative.
Page 25
2 Position the Miranti lift bath trolley over the bathtub and lower it down. (See Fig. 2) 3 Clean: • the upper part of the Miranti lift bath trolley (start from top and move downwards) • all removed parts from visible residue with a cloth soaked in water.
Page 26
7 Spray disinfectant on (See Fig. 3, 4 and 5): • the upper part of the Miranti lift bath trolley • all removed parts Use a brush soaked in disinfectant and scrub thoroughly, especially on handles and hand control.
Page 27
• the upper part of the Miranti lift bath trolley Fig. 6 • all removed parts • all other areas of the Miranti lift bath trolley, such as chassis The cloth should occasionally be rinsed in running water during removal of disinfectant. (See Fig. 6) 16 If all disinfectant cannot be removed, spray water on the affected part and wipe off with disposable towels.
Battery Instructions Installation of Charger WARNING See the NDA 1200, 2200, 4200, 6200 Battery Charger To avoid bodily injury, do NOT crush, IFU. puncture, open, dismantle or otherwise mechanically interfere with the battery. How to Change the Battery • Should the battery casing crack and See the NDA 1200, 2200, 4200, 6200 Battery Charger cause contents to come in contact with IFU.
Care and Preventative Maintenance The Miranti lift bath trolley is subject to wear and tear, and the following actions must be performed when specified to make sure that the product remains within its original manufacturing specification WARNING To avoid malfunction resulting in injury, make sure to conduct regular inspections and follow the recommended maintenance schedule.
Page 30
Miranti lift bath trolley knowledge following the instructions in this IFU. Every day • Disinfect. The Miranti lift bath trolley has to be disinfected immediately after every use. ArjoHuntleigh disinfectants and shampoos should be used in recommended concentrations. See section Cleaning and Disinfection Instructions on page 24.
Page 31
Check attachment and locking of side guard and patient security grip. All attachments shall shape-in firmly into all positions. (See Fig 8). Yearly The Miranti lift bath trolley has to be serviced according to the table QUALIFIED PERSONNEL Action/Check on page 32.
Page 32
WARNING To avoid injury and/or unsafe product, the maintenance activities must be carried out at the correct frequency by qualified personnel using correct tools, parts and knowledge of procedure. Qualified personnel must have documented training in maintenance of this device. QUALIFIED PERSONNEL Action/Check Every YEAR Every 2:nd YEAR...
In case of this scenario stop using the equipment and contact qualified personnel for repair of the equipment. NOTE If the product does not work as intended, immediately contact your local ArjoHuntleigh representative for support.
Fuse F1 F10AL 250 V Fuse F2 T2.5AL 250 V Allowed combinations The Miranti lift bath trolley panels integrate with Generic hospital/care facility bed with a the following ArjoHuntleigh products clearance of 150 mm (5 7/8”) System 2000 baths No other combinations are allowed.
Page 35
This requirement applies in the European Union. Do not dispose these products as unsorted municipal waste. You can return your device and accessories to ArjoHuntleigh or you can contact your local authorities for advice on disposal.
Labels Label explanation Separate electrical and electronic components for recycling in Attention Important user information for accordance with the European Label correct use. Directive 2002/96/EC (WEEE) Battery Label States safety and environmental Classified by Underwriters information for the battery. Laboratories Inc. with respect to Data Label States technical performance and electrical shock, fire, mechanical,...
EMC information below. The Miranti lift bath trolley is intended for use in the electromagnetic environment specified below. The customer or the user of the Miranti lift bath trolley should assure that it is used in such an environment.
Page 41
RF communications equipment (transmitters) and the Miranti lift bath trolley by maintaining a distance of not less than 1 meter to any part of the Miranti lift bath trolley, including cables, if the transmitters output power rating exceeds 1 W.
Parts and Accessories Electronic scale Safety belt For disinfectant contact CDA2100-01 8664559 the ArjoHuntleigh local representative Battery and battery Head rest Thigh support cushion charger Power drive...
Free: +1 800 072 040 Fax: +33 (0) 3 20 28 13 14 Fax: + 61 89337 9077 E-mail: info.france@arjohuntleigh.com POLSKA ArjoHuntleigh Polska Sp. z o.o. BELGIQUE / BELGIË HONG KONG ul. Ks Piotra Wawrzyniaka 2 ArjoHuntleigh NV/SA ArjoHuntleigh (Hong Kong) Ltd...
Page 44
Getinge Group is a leading global provider of products and systems that contribute to quality enhancement and cost effi ciency within healthcare and life sciences. We operate under the three brands of ArjoHuntleigh, Getinge and Maquet. Getinge provides solutions for infection control within healthcare and contamination prevention within life sciences. Maquet specializes in solutions, therapies and products for surgical interventions, interventional cardiology and intensive care.
Need help?
Do you have a question about the Miranti and is the answer not in the manual?
Questions and answers