5. Once patient surface has reached 0° supine:
Push in lock pin to stabilize patient surface. Ensure lock pin is fully engaged.
Re-engage emergency release handle to secure patient surface.
6. Unbuckle and stow chest pack.
7. Begin CPR as quickly as possible.
8. Unbuckle and stow remaining proning packs, as necessary.
9. Unfasten abdominal sling and fold to one side.
10. Remove the following, if necessary:
Arms from arm sling
Side support packs and accessory packs
Face pack
11. Place backboard under patient, following institution protocols and universal precautions.
12. After CPR is performed and patient is clinically stable, follow all applicable steps for securing patient and
reengaging therapy.
In case of stuck or jammed lock pin, refer to the trouble shooting section for stuck or
jammed lock pin instructions, page 2-27.
If the patient surface has been rotated 360° during the
emergency release, the information screen shown at right
will appear when power is restored to the RotoProne
unit. Press the Rotate 360° button as shown at right,
to return the patient surface to supine in the proper
direction. Monitor all tubes and lines as patient surface
returns to the supine position. Ensure patient is secure.