Arjohuntleigh RotoProne User Manual page 28

Table of Contents

Advertisement

Continue pressing either button until patient
surface is leveled, fully lowered and rotated into the
0° supine position.
Bed will stop at 67° if unsafe operating condition
occurs.
Once patient surface has reached 0° supine, the
CPR Information screen will appear.
3. Push in lock pin to secure surface.
Ensure lock pin is fully engaged.
4. Unbuckle and stow chest pack.
5. Begin CPR as quickly as possible.
6. Unbuckle and stow remaining proning packs, as
necessary.
7. Unfasten abdominal sling and fold to one side.
8. Remove the following, if necessary:
Arms from arm sling
Side support packs and accessory packs
Face pack
9. Place backboard under patient, following institution protocols and universal precautions.
10. After CPR is performed and patient is clinically stable, follow all applicable steps for securing patient and
reengaging therapy.