Power-Down for Shipping

1. Discontinue therapy.
2. Ensure patient surface is at 0° supine.
3. Push lock pin in.
4. Transfer patient from unit.
See Patient Care chapter of this manual for detailed instructions on patient transfer.
5. Press and hold OFF button on Home screen.
Shutdown Information screen will appear and alarm
will sound.
Alarm will sound until one of the buttons on the
screen is pressed.
All settings will be held in memory and restored
when the unit is plugged in.
6. Press Shutdown button (on the right side of the
screen). If the patient surface is not already leveled and
lowered, the Lock Pin Out screen will appear.
If the patient surface is leveled and lowered and
the lock pin is already pushed in, the Raise Unit
Information screen (described as follows in Step 9) will
appear.
7. Pull lock pin. Rotate and Lower Information screen
will appear.
8. Press and hold Rotate and Lower button until patient
surface is level at 0° supine and completely lowered.
Lock Pin In screen will appear.