7. Adjust (if necessary) head support alignment:
Assess position of patient's head and neck. Check for
proper alignment.
If alignment adjustment is required, turn head
support adjustment handle to raise or lower patient's
head to the desired position, as shown at right.
8. Reconfirm checklist. Once checklist is cleared, the Start Rotation screen will appear.
9. Press Start Rotation button. Prone Therapy will begin as programmed.
Push in lock pin to stop rotation at any time.
Once the patient is in the prone position, immediately open the upper body hatches to alleviate chest
compression and for better visibility and access to the patient.
Place surface in Reverse Trendelenburg to reduce facial pressure, abdominal pressure on diaphragm
and to feed. Ensure all lines and tubes remain attached as the surface moves into position.
10. Observe patient through two full rotation cycles and check the following:
Verify all lines and tubes are secured and have adequate slack to avoid pulling, binding or dislodging
during a full 360° of rotation. Adjust slack in lines, as necessary.
Re-tighten side support packs as necessary. Observe patient for movement or shift within packs. Pad
or adjust packs, as necessary.
Hatches should be opened while patient is in prone position for temperature control and patient
assessment.
Ensure there are no potential obstructions in path of unit rotation.
11. Re-examine patient's placement in head support assembly:
Re-assess position of face pack. Ensure patient's eyes are visible and there is no pressure on or around
eyes. Check regularly, as patient may migrate.
Re-assess patient's head and neck alignment and adjust head support up or down using head
adjustment handle, as needed.