Manual Rotation To Prone Instructions - Arjohuntleigh RotoProne User Manual

Table of Contents

Advertisement

Manual Rotation to Prone Instructions

Patient surface can be rotated without power to the prone position, if necessary. Use these emergency
release procedures to manually rotate the patient surface to prone position.
• At least three people recommended for task
• Patient must start at 0˚ supine
• Lock Pin must be pulled out
In case of stuck or jammed lock pin, refer to the trouble shooting section for stuck or
jammed lock pin instructions, page 2-27.
If patient is NOT 0˚ supine:
1. Instruct person 2 and 3 to secure frame at head-end and provide notification when frame is secure.
2. Disengage Emergency Release Handle.
Unit may shift when Emergency Release Handle is
disengaged.
3. Instruct person 2 and 3 to rotate patient to 0˚ supine.
When patient is 0˚ supine:
1. Determine proper direction for rotation.
2. Instruct person 2 and 3 to secure patient (including buckles and
hatches) for proning and monitor all tubes and lines.
3. Disengage emergency release handle,if necessary.
4. Break tab to open door and access manual rotation lever.
5. Lift and hold manual rotation lever for duration of manual rotation.
Audible notification will sound while manual rotation lever is active.
6. Instruct person 2 and 3 to rotate patient to prone position.
7. Release manual rotation lever and re-engage emergency release handle.
Call ArjoHuntleigh at 1-800-343-0974.
3
Manual
Rotation
Lever
Emergency
Release Handle
2-19
2
1

Hide quick links: