2. Pull lock pin. The Checklist screen will appear.
3. Check and clear each item on the checklist:
To clear each item, press the corresponding button.
The arrow will change to a check mark.
Assess all invasive lines and tubes. Ensure lines are
secure and have adequate slack to avoid pulling,
tangling, binding or dislodging during a full 360° of
rotation. Once all lines are assessed to be adequately
secured, press Check Tubing button.
Examine patient's artificial airway system. Ensure ETT and ventilator tubing are secure and have
adequate slack to avoid pulling or twisting during a full 360° of rotation. Once the airway is assessed
to be adequately secured, press Check Airway button.
Consider labeling emergency medication line for rapid access.
Examine patient's placement in head support assembly. Ensure patient's ears are aligned with
openings in the sides of head support assembly. Ensure patient's eyebrows are aligned with top of
opening in face pack. Ensure that face pack (if used) is positioned so that patient's eyes are visible
and no part of pack places pressure on or around the eye and orbit area. Ensure face pack straps are
secure and tightened appropriately. Once patient's head is cleared, press Check Head Support
button.
Tightness of straps will vary according to each patient's needs. Secure straps snugly to minimize
patient's shift away from patient surface when moved into a prone position.
Once checklist has been cleared, the Start
Rotation screen will appear.
3-24